error ratio for k type
This commit is contained in:
parent
839da93a3e
commit
676838815b
@ -573,7 +573,7 @@ but statistically less reliable.
|
|||||||
|
|
||||||
Because the potential divider provides an offset as a side effect of detecting a disconnection
|
Because the potential divider provides an offset as a side effect of detecting a disconnection
|
||||||
resistance in the thermocopule extension or compensation cable will have an effect.
|
resistance in the thermocopule extension or compensation cable will have an effect.
|
||||||
For a `k' type thermocouple this would be of the order of $5 { }^{o}C$ for $100\Omega$ of
|
For a `k' type thermocouple this would be of the order of $0.5 { }^{o}C$ for $10\Omega$ of
|
||||||
cable loop impedance. Therefore, accuracy constraints and cable impedance should be considered
|
cable loop impedance. Therefore, accuracy constraints and cable impedance should be considered
|
||||||
to determine specified maximum compensation/extension lengths.
|
to determine specified maximum compensation/extension lengths.
|
||||||
|
|
||||||
|
Loading…
Reference in New Issue
Block a user