@ -573,7 +573,7 @@ but statistically less reliable.
|
||||
|
||||
Because the potential divider provides an offset as a side effect of detecting a disconnection
|
||||
resistance in the thermocopule extension or compensation cable will have an effect.
|
||||
For a `k' type thermocouple this would be of the order of $5 { }^{o}C$ for $100\Omega$ of
|
||||
For a `k' type thermocouple this would be of the order of $0.5 { }^{o}C$ for $10\Omega$ of
|
||||
cable loop impedance. Therefore, accuracy constraints and cable impedance should be considered
|
||||
to determine specified maximum compensation/extension lengths.
|
||||
|
||||
|